| Model | CALsys 1200/Autocal |
|---|---|
| Temp Range | 250 °C ~ 1200 °C |
| Stability | ±0.1°C at 250°C ±0.2°C at 700°C ±0.3°C at 1200°C |
| Radial Uniformity | ±0.20°C at 250°C ±0.24°C at 700°C ±0.36°C at 1200°C |
| Immersion depth | 160mm |
| Insert OD Dimensions | 37mm |
| Method of Control | Self Tunned PID Controller |
| Heating time | 60 min |
| Cooling time | 150 min(1200⁰C ~ 250⁰C ) |
| Resolution | 0.1⁰C up to 999⁰C |
| Computer interface | RS- 232 |
| Input (CALsys 1200 Autocal) | Four Channels (One master and three test sensors), high-quality LEMO connector suitable for both T/C (J, K, N, T, R, S, B type) and Rtd. |
| Software (CALsys 1200 Autocal) | The calibrator should be provided with software for data recording (Manual Mode) and Test Certificate generation in Auto Mode |
| Data logging (CALsys 1200 Autocal) | Data logging facility with logged data export to computer through LAN port (optional USB) |